• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SAS ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SAS ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA ONE 3.1
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2022
Event Type  Injury  
Event Description
The screen of the robot did not turn on, so it was not usable for a brain biopsy.
 
Manufacturer Narrative
Tests were done remotely to solve this issue.There was no display on the screen even though it was well powered (power light on).The surgeon restarted the system, this time by connecting the camera stand and there the display on the screen reappeared.Finally the surgeon restarted the rosa robot alone, without the camera connected, and everything worked.The most probable root cause of this event would be an electrical issue, but this cannot be confirmed.Unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SAS
432 rue du rajol
mauguio, languedoc-roussillon 34130
FR  34130
Manufacturer (Section G)
MEDTECH SAS
432 rue du rajol
mauguio, languedoc-roussillon 34130
FR   34130
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key13540129
MDR Text Key288026352
Report Number3009185973-2022-00011
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K192173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA ONE 3.1
Device Catalogue NumberROSAS00201
Device Lot Number3.1.5.28
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2022
Initial Date FDA Received02/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-