A haemonetics field service engineer evaluated the nexsys pcs300 system used in the procedure.The system log files were downloaded, the device was fully tested and a water test was performed.There were no issues found with the equipment; testing showed no malfunction.The status of the disposables used in the procedure is unknown; without the samples, haemonetics is unable to perform further evaluation.
|
On (b)(6), 2022, haemonetics was notified of the following during a donation procedure in belgium, utilizing a nexsys pcs 300 system: the injection went smoothly, samples were taken well.Smooth start of the purchase.After a few minutes an alarm came up on the screen that the plasma flow was not detected.No kinks in the line.Blood cells returned.Then always alarm that air was detected and collection stopped.During donation.Actions already taken.Decrease stopped.Donor slightly vagal, probably due to emotion, slight reaction to citrate, ca tablet given.
|