MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION HALL EASY FIT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION HALL EASY FIT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number A7700E20
Device Problems Insufficient Information (3190); Patient Device Interaction Problem (4001)
Patient Problems Foreign Body Reaction (1868); Thrombosis/Thrombus (4440); Aortic Valve Insufficiency/ Regurgitation (4450); Insufficient Information (4580)
Event Date 12/20/2021
Event Type  Injury  
Event Description
Medtronic received information that 17 years and 8 months post implant of this 20mm aortic mechanical valve, it was explanted and re placed with a 21mm bioprosthetic valve.The reason for the replacement was not reported.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic received additional information that the reason for replacement was due to valve thrombus and mild aortic regurgitation.It was reported that during the explant procedure, pannus growth was noted on both the right and left leaflets that required a fair amount of force to open the leaflets.No additional adverse patient effects were reported.A4: patients weight updated h6: codes updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALL EASY FIT
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13545890
MDR Text Key285755411
Report Number2025587-2022-00464
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P790018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA7700E20
Device Catalogue NumberA7700E20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2022
Initial Date FDA Received02/16/2022
Supplement Dates Manufacturer Received02/28/2022
Supplement Dates FDA Received03/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age68 YR
Patient SexFemale
Patient Weight58 KG
-
-