Model Number N/A |
Device Problems
Fracture (1260); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/25/2022 |
Event Type
malfunction
|
Event Description
|
It was reported the screwdriver tip broke when tightening the set screw in the trial.The tip become engaged and was not able to extract the screwdriver tip after many different attempts.Surgeon had to take out the entire construct and waste the stem.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Product is in process of being returned.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The hex driver was not returned but the associated femoral stem was returned.No part of the hex driver was found in the stem.Visual and dimensional evaluations could not be performed for the driver.Lot identification is necessary for review of device history records; lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information on the reported event.
|
|