Model Number CDI510H
Device Problem Insufficient Information (3190)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that there was a product complaint.It is unknown when this event occurred, whether the product was changed out, or if there was any effect on the patient or results of the surgery.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 17, 2022.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 4114, 3221, 4315).Type of investigation: 4114 - device not returned.Investigation findings: 3221 - no findings available.Investigation conclusions: 4315 - cause not established.An investigation could not be performed and a root cause could not be determined due to insufficient description details, the lot number not being provided and no samples being returned.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.  upon further investigation of the reported event, the following information is new and/or changed: b5 (updated describe event or problem).D4 (additional device information - added lot number).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).A third follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
Additional information received that states, the product to freeze during transport because it was not packaged in the gel box against the cold.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H4 (device manufacture date).The affected box of shunt sensors was not returned and photos were not provided.The reported lot was shipped to canada; therefore, gel packs should have been packed in the over pack box per procedure.A thorough investigation could not be performed and a definitive root cause could be determined.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key13548413
MDR Text Key285740053
Report Number1124841-2022-00024
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00699753160767
UDI-Public(01)00699753160767
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/31/2022
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberZN29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/02/2022
Initial Date FDA Received02/17/2022
Supplement Dates Manufacturer Received04/13/2022
06/16/2022
09/16/2022
Supplement Dates FDA Received04/18/2022
07/07/2022
09/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-