MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-401L-0523
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information obtained stated the customer discarded the product.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
It was reported to olympus that during an unspecified therapeutic procedure the fluid would not flow through the single-use injector needle when advanced outside of the sheath.When retracted into the sheath, fluid flowed through the needle.The procedure was completed using a similar device.There was no report of patient death, harm, or injuries.The following patient identifiers are related events: (b)(6).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.The device lot number was not received however, the device history records (dhrs) for the past year prior to the aware date of this event were inspected and no abnormalities were found that could have led to the reported event.The age of the device is unknown since the lot number was not received.Based on the results of the investigation, the exact cause could not be conclusively identified since the device was not retuned.However, based on a similar complaint in the past, it is likely that the phenomenon occurred due to the compressive bucking on the needle tube.It is unlikely a unit with defective needle is shipped as nm-401l series undergo 100% inspection for appearance, needle operation and injection.The compressive buckling on the needle tube was likely caused when the needle was extended because of the great friction between the outer tube and the needle.It was likely that the friction between the outer tube and the needle increased by the following factors ·the needle extended/retracted while the tube was coiled in inspection of operation.·the slider was abruptly pushed.·the kink of the tube.·angle of the distal end of the endoscope.The instruction manual contains the following descriptions, and it warns against this event.(rk1556 rev.01).·straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.·operate the slider slowly, otherwise the tube could buckle.·when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.·insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.·stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.Although the actual defect was not sent, it is possible to infer the cause from the investigation results of similar cases in the past, so it is unnecessary to investigate using the same structure of equipment and similar equipment.It was judged.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Type of Device
INJECTOR AND SHEATHSET
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13552142
MDR Text Key296382942
Report Number8010047-2022-03150
Device Sequence Number1
Product Code FBK
UDI-Device Identifier04953170422645
UDI-Public04953170422645
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K902736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNM-401L-0523
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2022
Initial Date FDA Received02/17/2022
Supplement Dates Manufacturer Received05/30/2022
Supplement Dates FDA Received06/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-