Model Number 102096-67A |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/05/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
This mdr is result of a retrospective review of complaints.The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
|
Event Description
|
On 30 march 2021, senseonics was made aware of an incident where user complains of having low signal within the transmitter and sensor, leading to early sensor removal as the patient was having issues with placement guide weak connection.
|
|
Manufacturer Narrative
|
The customer complaint could not be confirmed as sensor communicated throughout the in-vitro qc test as well as during the two other above-described tests.The potential root cause of the inability of the customer to get a "good/excellent" signal is that the sensor was too deeply inserted in her arm as could be read in the case notes.This case was initially dispositioned as non-complaint and re-dispositioned as complaint per ntf-0456.As part of resolution, the rma was authorized to offer the user a sensor replacement.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 213.H6.Investigation conclusions updated to 67.
|
|