• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TAEWOONG MEDICAL CO.,LTD. NITI-S ESOPHAGEAL COVERED STENT; ESOPHAGEAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TAEWOONG MEDICAL CO.,LTD. NITI-S ESOPHAGEAL COVERED STENT; ESOPHAGEAL STENT Back to Search Results
Model Number EA2210F
Device Problems Migration or Expulsion of Device (1395); Reflux within Device (1522); Migration (4003)
Patient Problem Vomiting (2144)
Event Date 01/25/2022
Event Type  Injury  
Event Description
Patient complaint about sever reflux and started vomiting.Patient was on double ppi and food intake was monitored; barium swallow was 100%.Dr scoped her and saw that the valve was inverted and stuck to the stent, he tried to push the valve down but couldn't manage and the stent migrated.There was a problem when we tried to remove the delivery stent on placing the stent, it wanted to pull out the stent with the delivery system.We managed to get the delivery system out without removing the stent itself.We have closed the delivery system before removing it.
 
Manufacturer Narrative
It was reported that during removal of the delivery system after placement of the stent, the skirt was caught in the delivery system but the physician managed to remove the delivery system.Patient complained about reflux and vomit.When checking with the scope, the valve was inverted, and when he tried to push the valve, the stent migrated.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.Based on the description "saw that the valve was inverted and stuck to the stent, he tried to push the valve down but couldn't manage and the stent migrated" and "there was a problem when we tried to remove the delivery stent on placing the stent, it wanted to pull out the stent with the delivery system", it is considered the tip of the delivery system was stuck to the skirt when removing the delivery system after stent placement.It is considered this caused the stent to be blocked, and the patient to complain of reflux and vomit, and when the physician tried to push the skirt, the stent migrated and was removed.Through the user manual by taewoong, it is stated that "after stent deployment, removal of the introducer system and guide wire from the patient should be handled with care.During removal, if strong resistance is felt, wait 3-5 minutes for the stent to expand then try again" and "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: reflux" this suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ESOPHAGEAL COVERED STENT
Type of Device
ESOPHAGEAL STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, 10022
KS   10022
Manufacturer Contact
lee
14, gojeong-ro
wolgot-myeon
gimpo-si, 10022
MDR Report Key13574774
MDR Text Key290402430
Report Number3003902943-2022-00006
Device Sequence Number1
Product Code ESW
Combination Product (y/n)Y
Reporter Country CodeSF
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/01/2024
Device Model NumberEA2210F
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/01/2022
Initial Date FDA Received02/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-