Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device would not power on.No patient injury was reported.
 
Event Description
It was reported that the device would not power on.No patient injury was reported.
 
Manufacturer Narrative
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device presented a cracked enclosure, a faulty pcb, wear and tear damaged line cord, pole clamp, and tank cover, outdated pcb and power switch.The customer stated problem was duplicated.There was no power so a test pcb and power switch were installed.Plugged the line cord into a power source, turned on the power switch, and the device turned on.The connection between the power switch and pcb was damaged.Replaced: enclosure, tank cover, line cord, and pole clamp.Also upgraded the damaged pcb and power switch.The cause of the reported problem was wear and tear from customer repeated use of the power switch.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13584305
MDR Text Key286006991
Report Number3012307300-2022-03914
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2022
Initial Date FDA Received02/23/2022
Supplement Dates Manufacturer Received05/19/2022
Supplement Dates FDA Received05/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2014
Is the Device Single Use? No