H3 : product analysis : further analysis of bur found that only the tang and portions of the shaft were returned.Visually, there was no damage to the construction of the device that would have resulted in the reported event.The length of the tang should measure 0.750 inches and measured 0.75 inches which was within specification.The distal outside diameter or the tang should measure 0.0919 ± 0.003 inches and measured 0.092 inches which was within specification.The proximal outside diameter of the tang should measure 0.056 ± 0.002 inches and measured 0.055 inches which was within specification.The other portions of the bur assembly were not adhered to the tang which include the bur wire, the housing, and tubing.Functional testing could not be performed due to the missing portion of the device.In the returned condition, there was an out of specification condition that was related to the complaint due to physical damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|