• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® BUR - UNKNOWN; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® BUR - UNKNOWN; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number XOM UNK BUR
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2022
Event Type  malfunction  
Manufacturer Narrative
Analysis found that the handpiece will not lock due to a bur broken off and stuck in the bur lock.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Service and repair found a bur that was broken off and stuck in the handpiece.
 
Manufacturer Narrative
Analysis found that the handpiece will not lock due to a bur broken off and stuck in the bur lock.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Service and repair found a bur that was broken off and stuck in the handpiece.
 
Manufacturer Narrative
H3 : product analysis : further analysis of bur found that only the tang and portions of the shaft were returned.Visually, there was no damage to the construction of the device that would have resulted in the reported event.The length of the tang should measure 0.750 inches and measured 0.75 inches which was within specification.The distal outside diameter or the tang should measure 0.0919 ± 0.003 inches and measured 0.092 inches which was within specification.The proximal outside diameter of the tang should measure 0.056 ± 0.002 inches and measured 0.055 inches which was within specification.The other portions of the bur assembly were not adhered to the tang which include the bur wire, the housing, and tubing.Functional testing could not be performed due to the missing portion of the device.In the returned condition, there was an out of specification condition that was related to the complaint due to physical damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR - UNKNOWN
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key13585254
MDR Text Key286241989
Report Number1045254-2022-00092
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberXOM UNK BUR
Device Catalogue NumberXOM UNK BUR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2022
Initial Date FDA Received02/23/2022
Supplement Dates Manufacturer Received02/23/2022
Supplement Dates FDA Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-