• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 21AECJ-502
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/30/2021
Event Type  Injury  
Event Description
It was reported, a 21mm sjm masters series valve expanded cuff was chosen for procedure.During the procedure while the user was positioning the device and rotated the valve, it was reported that the user applied increased tension to the valve as the it was not rotating properly and the leaflet broke.The device was removed and exchanged with a new 21mm sjm masters series valve expanded cuff.No additional information has been provided.The event did cause a reported extra 45 minutes on cardiopulmonary bypass and was stable throughout the procedure.No additional information has been provided.
 
Manufacturer Narrative
The reported event of a "broken" leaflet was confirmed.Both leaflets was dislodged from the orifice and returned with the valve.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment could not be conclusively determined, however information from the field indicated that the user applied additional tension to the valve, which could have contributed to the dislodged leaflets.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
177 east county road b
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
177 east county road b
st. paul MN 55117
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13587924
MDR Text Key286051016
Report Number3005334138-2022-00143
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734053884
UDI-Public05414734053884
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/07/2022
Device Model Number21AECJ-502
Device Catalogue Number21AECJ-502
Device Lot Number6128571
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/01/2022
Initial Date FDA Received02/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient Weight69 KG
-
-