Model Number N/A |
Device Problems
Positioning Failure (1158); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: 42532007101 natural tibia cemented 5 degree stemmed left size e - 65105634.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the poly would not lock into the tibial tray and upon inspection, the underside of the poly appeared dented.The surgical technique was utilized.There was no surgical delay.No known impact or consequence to the patient.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified damage (nicked/gouged) on the distal surface of implant.The locking features was found to be compressed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented before pushing it using the inserter instrument.Detailed instructions for inserting the articular surface are provided in the surgical technique.The root cause is attributed to use error.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|