• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORI; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORI; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 402810
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2022
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During an atrioventricular nodal reentrant tachycardia procedure, due to a noise issue and the troubleshooting involved, a prolonged procedure occurred.Initially the cable and generator were exchanged, but the issue remained.The catheter was then exchanged and the issue was resolved.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORI
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13599005
MDR Text Key287141646
Report Number2182269-2022-00014
Device Sequence Number1
Product Code LPB
UDI-Device Identifier05414734206013
UDI-Public05414734206013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2024
Device Model Number402810
Device Catalogue Number402810
Device Lot Number7996499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received02/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-