A us customer observed elevated atellica im br 27.29 (br) results from a sample that was considered discordant with the replicate testing.Siemens is investigating.The limitations section of the instructions for use states the following: "note do not interpret levels of ca 27.29 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed breast carcinoma frequently have levels of ca 27.29 within the range observed in healthy individuals.Additionally, elevated levels of ca 27.29 can be observed in patients with nonmalignant diseases.Measurements of ca 27.29 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation.".
|
The initial mdr 1219913-2022-00069 was filed on february 25, 2022.Additional information - march 08, 2022.A us customer observed a sample that initially recovered 45 u/ml with atellica im br 27.29 (br) assay kit lot 261.When the sample was repeated the next day on the same analyzer it recovered 144 u/ml.Further repeat testing of the same sample for 4 more times on the same analyzer 2 hours later it recovered 55.4, 40.5, 39.1, and 46.9 u/ml.The 144 u/ml result was an outlier but the other values were within the expected performance for the assay.In the initial mdr, the result of 55.4 u/ml was identified conservatively as a discordant result.A review of data indicated the issue was not due to the reagents or the calibration.A review of the reagent probe trace files did not show any issue with reagent addition.A review of the sample traces files suggested that there was particulate matter in the sample which could have interfered with the atellica im br 27.29 (br) assay and led to the elevated result of 144 u/ml.The cause of the 144 u/ml result when using atellica im br 27.29 (br) assay kit lot 261 could not be determined but siemens cannot rule out preanalytical factors or a sample issue.Based on the investigation, no product problem was identified.The customer is operational.No further action is required.In section h6, investigation findings and investigation conclusion codes were updated based on the investigation results.
|