Model Number CLV-190
Device Problems Corroded (1131); Moisture Damage (1405)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation revealed the base plate, housing cover, lamp door, front panel, scope socket, and capacitor had corrosion.Additionally, a non-olympus lamp was installed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus the device had circuit board damage.Upon inspection and testing of the customer returned device, it was observed that the device had moisture penetrating various housing and parts and showed signs of corrosion.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The device evaluation revealed the base plate, housing cover, lamp door, front panel, scope socket, and capacitor had corrosion.Additionally, a non-olympus lamp was installed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus the device had circuit board damage.Upon inspection and testing of the customer returned device, it was observed that the device had moisture penetrating various housing and parts and showed signs of corrosion.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the flooding could not be determined.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿[7.1 care] do not clean the output socket, other connectors, or the ac power inlet.Cleaning them can deform or corrode the contacts resulting in damage to the light source.Do not immerse, autoclave, or gas sterilize the light source.These methods will damage it.[6.1 replacement of the examination (xenon) lamp] never install a lamp that has not been approved by olympus.The use of a nonapproved lamp can cause damage to the light source and ancillary equipment, malfunction or a fire.¿olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13602036
MDR Text Key286144755
Report Number8010047-2022-03585
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/31/2022
Initial Date FDA Received02/25/2022
Supplement Dates Manufacturer Received05/13/2022
Supplement Dates FDA Received06/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-