• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 UNKNOWN LATEX TEMPERATURE SENSING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 UNKNOWN LATEX TEMPERATURE SENSING CATHETER Back to Search Results
Device Problem Inaccurate Flow Rate (1249)
Patient Problem Urinary Retention (2119)
Event Date 02/02/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the foley temperature sensing catheters were not draining appropriately and resulted in significant retention in two patients.Customer stated that they were diligent about positioning the catheter to avoid kinks or occlusions.One patient had statlock in place, but the other did not have it.Customer mentioned that it was not just a little retention.One of the patients was consulted renal for an elevated creatinine.There were no obvious defects found when the catheter was exchanged and the catheter itself looked patent.The second patient had some significant chugging or elevated venous drainage pressures (ecmo) related to bladder distention.It was further reported that few other nurses have faced similar issues where they had to exchange the foley catheter and 2 liters of urine drained.This also happened last night for retention issues with bard foley catheter.No medical intervention was reported.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the foley temperature sensing catheters were not draining appropriately and resulted in significant retention in two patients.Customer stated that they were diligent about positioning the catheter to avoid kinks or occlusions.One patient had statlock in place, but the other did not have it.Customer mentioned that it was not just a little retention.One of the patients was consulted renal for an elevated creatinine.There were no obvious defects found when the catheter was exchanged and the catheter itself looked patent.The second patient had some significant chugging or elevated venous drainage pressures (ecmo) related to bladder distention.It was further reported that few other nurses have faced similar issues where they had to exchange the foley catheter and 2 liters of urine drained.This also happened last night for retention issues with bard foley catheter.No medical intervention was reported.
 
Event Description
It was reported that the foley temperature sensing catheters were not draining appropriately and resulted in significant retention in two patients.Customer stated that they were diligent about positioning the catheter to avoid kinks or occlusions.One patient had statlock in place, but the other did not have it.Customer mentioned that it was not just a little retention.One of the patients was consulted renal for an elevated creatinine.There were no obvious defects found when the catheter was exchanged and the catheter itself looked patent.The second patient had some significant chugging or elevated venous drainage pressures (ecmo) related to bladder distention.It was further reported that few other nurses have faced similar issues where they had to exchange the foley catheter and 2 liters of urine drained.This also happened last night for retention issues with bard foley catheter.No medical intervention was reported.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿inadequate material selection".It was unknown whether the device had met relevant specifications.The product was used for diagnostics and treatment purposes.It was unknown whether the product had caused the reported failure.The lot number was unknown; therefore, the device history record could not be reviewed.The product catalog number and the lot number for this device was unknown.Therefore, bd is unable to determine the associated labeling to review.The device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LATEX TEMPERATURE SENSING CATHETER
Type of Device
UNKNOWN LATEX TEMPERATURE SENSING CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13604085
MDR Text Key286337114
Report Number1018233-2022-00885
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/02/2022
Initial Date FDA Received02/25/2022
Supplement Dates Manufacturer Received04/16/2022
Supplement Dates FDA Received05/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-