• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794001
Device Problems Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2022
Event Type  malfunction  
Manufacturer Narrative
The meter was requested for investigation.
 
Event Description
The initial reporter stated there was an issue with the display of a coaguchek xs meter.The meter was powered on and had segments missing from the results field of the display.A display check of the meter was performed and only the tops of the "888" were seen in the results field.When checking the meter memory, a result of 19.1 seconds (41 % quick, 1.6 inr) was observed.
 
Manufacturer Narrative
The meter was returned for investigation.A display test showed no faded or missing segments.When displaying measurement results, some segments are displayed with poor contrast.The circuit board was tested for damage or contamination.The battery compartment and circuit board showed contamination by liquid (leaked battery) which has penetrated/corroded the solder contacts.The contamination also affected the conductive rubber contacts, causing the low contrast display.The contamination issue would be related to a handling error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13606269
MDR Text Key289063085
Report Number1823260-2022-00524
Device Sequence Number1
Product Code GJS
UDI-Device Identifier04015630945689
UDI-Public04015630945689
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/03/2022
Initial Date FDA Received02/25/2022
Supplement Dates Manufacturer Received03/24/2022
Supplement Dates FDA Received04/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-