• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARTIVA CARTIVA; PROSTESIS, METATARSOPHALANGEAL JOINT CARTILAGE REPLACEMENT IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARTIVA CARTIVA; PROSTESIS, METATARSOPHALANGEAL JOINT CARTILAGE REPLACEMENT IMPLANT Back to Search Results
Model Number CAR-10
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 04/30/2018
Event Type  Injury  
Manufacturer Narrative
The product was returned for review.The height of the device and diameter were within the post-sterilization value range.Visual examination found on the articular surface sucffing, a smooth cut as from a scalpel blade, and a rough depression on the radiused edge of the articulating surface.The depression approximated the shape of the 2mm guide pin provided with the cartiva instrumentation.The scuffing is consistent with scuffing observed in the cartiva on cartilage wear study.More detail is available on the attached report.On november 10th, 2020 stryker trauma and extremities acquired cartiva, wright medical, tornier inc.And tornier sas.An effort was made to align the acquired entity complaint files with stryker's policies and procedures.Stryker has performed retrospective review of complaints and the reportability decision was revised to report this incident under the mdr/vigilance process.
 
Event Description
Explant performed on unidentified patent after two years implanted.No further details known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTIVA
Type of Device
PROSTESIS, METATARSOPHALANGEAL JOINT CARTILAGE REPLACEMENT IMPLANT
Manufacturer (Section D)
CARTIVA
6120 windward parkway
suite 220
alpharetta GA 30005
Manufacturer (Section G)
CARTIVA, INC
6120 windward parkway
suite 220
alpharetta GA 30005
Manufacturer Contact
matthew parrish
1023 cherry rd
memphis, TN 38117
MDR Report Key13606410
MDR Text Key286226769
Report Number3009351194-2022-00024
Device Sequence Number1
Product Code PNW
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P150017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2018
Device Model NumberCAR-10
Device Lot NumberF021116002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2018
Initial Date Manufacturer Received 05/01/2018
Initial Date FDA Received02/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age52 YR
Patient SexMale
-
-