The product was returned for review.The height of the device and diameter were within the post-sterilization value range.Visual examination found on the articular surface sucffing, a smooth cut as from a scalpel blade, and a rough depression on the radiused edge of the articulating surface.The depression approximated the shape of the 2mm guide pin provided with the cartiva instrumentation.The scuffing is consistent with scuffing observed in the cartiva on cartilage wear study.More detail is available on the attached report.On november 10th, 2020 stryker trauma and extremities acquired cartiva, wright medical, tornier inc.And tornier sas.An effort was made to align the acquired entity complaint files with stryker's policies and procedures.Stryker has performed retrospective review of complaints and the reportability decision was revised to report this incident under the mdr/vigilance process.
|