Model Number 1616000
Device Problem Inaccurate Information (4051)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 03/2023).
 
Event Description
It was reported that prior to a port placement procedure, upon scanning the barcode of the device the scanned product allegedly displays a different reference number instead of the original reference number.There was no patient contact.
 
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 03/2023).
 
Event Description
It was reported that prior to a port placement procedure, upon scanning the barcode of the device the scanned product allegedly displays a different reference number instead of the original reference number.There was no patient contact.
 
Event Description
It was reported that upon scanning the barcode of the device, the scanned product allegedly displays a different reference number instead of the original reference number.There was no patient involvement.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was not returned for evaluation, one electronic photo was provided for review.The photo shows a product packaging label of a powerport clearvue slim implantable port.The label indicates product catalog number 1616000, lot number refx4030 and expiration date 2023-03-31.A bar code on the bottom of the label was noted to be highlighted by the user.However, the provided photo is not sufficient enough to confirm the reported event.Therefore, the investigation is inconclusive for the reported inaccurate information.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 03/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT CLEARVUE SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13618369
MDR Text Key286246255
Report Number3006260740-2022-00528
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741026447
UDI-Public(01)00801741026447
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1616000
Device Catalogue Number1616000
Device Lot NumberREFX4030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/01/2022
Initial Date FDA Received02/28/2022
Supplement Dates Manufacturer Received05/10/2022
Supplement Dates FDA Received05/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-