SMITH & NEPHEW, INC. CONQUEST FN PROX LCK SCR DRILL GUIDE; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT
|
Back to Search Results |
|
Model Number 71179203 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that, during use, the threads of the conquest fn prox lck scr drill guide stripped and broke off.A back-up device was available.It is unknown if there was a delay due to this event.No patient complications reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|