• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CONQUEST FN PROX LCK SCR DRILL GUIDE; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CONQUEST FN PROX LCK SCR DRILL GUIDE; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT Back to Search Results
Model Number 71179203
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2022
Event Type  malfunction  
Event Description
It was reported that, during use, the threads of the conquest fn prox lck scr drill guide stripped and broke off.A back-up device was available.It is unknown if there was a delay due to this event.No patient complications reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST FN PROX LCK SCR DRILL GUIDE
Type of Device
DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13625971
MDR Text Key287285248
Report Number1020279-2022-00925
Device Sequence Number1
Product Code JDO
UDI-Device Identifier00885556533086
UDI-Public00885556533086
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152686
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71179203
Device Catalogue Number71179203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/08/2022
Initial Date FDA Received02/28/2022
Supplement Dates Manufacturer Received03/24/2022
Supplement Dates FDA Received03/27/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-