Lot Number GA0213
Device Problem Excessive Heating (4030)
Patient Problems Burn(s) (1757); Itching Sensation (1943); Pain (1994)
Event Date 01/30/2022
Event Type  Injury  
Manufacturer Narrative
The root cause cannot be identified.The site investigated this complaint by reviewing the device history records and manufacturing controls.The review of the device history records, batch thermal records, and production controls met the product release criteria.Consumer reports "she felt a burning and pain at the application site.She had red spots that looked like burns.No blisters and no open sores".The cause of the consumer stating she "she felt a burning and pain at the application site.She had red spots that looked like burns.No blisters and no open sores" is inconclusive since review of records does not provide evidence to support defective product.The product effect may vary with each individual.Care should be taken when using the device, following all safety and use information as provided with the wrap to avoid the risks of burns or other skin irritations.This is an adverse event for a burn.A risk calculation cannot be determined as there is no reasonable suggestion of a device malfunction.
 
Event Description
On 04-feb-2022, a spontaneous report from the united states was received from a consumer regarding a (b)(6) female who was using a thermcare neck/shoulder/wrist heat wrap.Additional information was received on 07-feb-2022.Medical history included hypertension and high cholesterol.The consumer had no known allergies.Concomitant products included hydrochlorothiazide and lipitor (atorvastatin).On (b)(6) 2022, the consumer topically applied a thermcare neck/shoulder/wrist heat wrap (expiration date: 30-sep-2024; lot number: ga0213) to her neck over her t-shirt.On (b)(6) 2022, approximately 5 minutes after application of the product, she felt a burning and pain at the application site.She took the heat wrap off and it looked red, like it burned her.She had a "pretty bad burn".Her daughter looked at her neck and noted she had red spots that looked like burns.There were no blisters and no open sores.Her daughter wanted her to go to her health care provider but she refused.For treatment she used neosporin and a band-aid.As of (b)(6) 2022, the area looked better but was still red and a little itchy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMACARE NECK/SHOULDER/WRIST
Type of Device
HOT OR COLD DISPOSABLE PACK
Manufacturer (Section D)
ANGELINI
1231 wyandotte dr
albany GA 31705
Manufacturer (Section G)
BRIDGES CONSUMER HEALTHCARE
811 broad street, suite 600
chattanooga TN 37402
Manufacturer Contact
scott hughes
811 broad street, suite 600
chattanooga, TN 37402
4237178579
MDR Report Key13635492
MDR Text Key286352713
Report Number3007593958-2022-00009
Device Sequence Number1
Product Code IMD
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot NumberGA0213
Initial Date Manufacturer Received 02/04/2022
Initial Date FDA Received03/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HYDROCHLOROTHIAZIDE; LIPITOR (ATORVASTATIN)
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight75 KG
-
-