Catalog Number CDS0705-NT |
Device Problems
Positioning Failure (1158); Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is being filed to report the cds could not straighten.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The clip delivery system (cds) was advanced into the left ventricle and then pulled back into the left atrium (la) because it was difficult to approach the target site.M knob was applied however the shaft did not straighten therefore the cds was removed.One clip was implanted, reducing mr to 1.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
The reported failure to advance could not be replicated in a testing environment as it could be related to patient anatomy/procedural operational circumstances.The steerable sleeve functioned as intended as the m should not be straighten.The curves were hold in place and no knob slippage was observed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lots and a review of the complaint history identified no other incident reported from this lot.Based on available information, a cause for the reported failure to advance issue could not be determined.The reported unable to straighten was a normal function of the device.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|