Catalog Number 383019
Device Problem Free or Unrestricted Flow (2945)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 01/26/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: in response to the event reported, a device history review was conducted for lot number 1293557.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
 
Event Description
It was reported that the bd intima ii¿ iv catheter prn adapter tubing clamp came loose and blood flowed back through the cap when it was opened.The following information was provided by the initial reporter, translated from (b)(6): "for this patient at the external jugular vein using this indwelling needle, after puncturing well to close the white stopcock, ready to replace the heparin cap, just open the original cap, blood immediately back out, pick up the hole syringe back to draw, without any resistance to draw back blood, the process of stopcock are closed, indicating that the indwelling needle stopcock did not play the effect, and later replaced the indwelling needle.No major impact on the patient, but increased risk of infection.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA II¿ IV CATHETER PRN ADAPTER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH  215126
Manufacturer (Section G)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH   215126
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13659728
MDR Text Key286960160
Report Number3014704491-2022-00085
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383019
Device Lot Number1293557
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/11/2022
Initial Date FDA Received03/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-