OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2022
Event Type  malfunction  
Manufacturer Narrative
During troubleshooting the customer was advised to verify all gauges around fifty (50) for all of them, select the rinse function the customer informed the service center representative all oer-pro gauges dropped to thirty (30).After the rinse cycle completed the customer confirmed the gauges is now at twenty-nine (29).Lastly the customer was instructed to purge the air from the internal water filter, by connecting tubing and selecting leak test to verify the time code is holding at twenty-nine (29).The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
It was reported to olympus that the after replacing the external water filters and the internal water filter on the oer-pro endoscope reprocessor, the time code on the cycle is thirty-one (31).There was no reports of patient death, harm, or injuries related to this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the water supply in the reprocessing basin was weaker than before because of inadequate air drainage from the water filter.To prevent recurrence, the instructions for use advise the user to ensure air drainage when replacing water filter.Also, low water supply from the main valve at the facility was another likely caused.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13666548
MDR Text Key296765399
Report Number8010047-2022-03945
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/07/2022
Initial Date FDA Received03/04/2022
Supplement Dates Manufacturer Received04/27/2022
Supplement Dates FDA Received05/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-