• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number PRT-01201-000
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2022
Event Type  malfunction  
Manufacturer Narrative
Ventec will perform an evaluation of the device.A follow-up report will be submitted when the investigation is complete as defined by 21 cfr 803.56.
 
Event Description
It was reported to ventec that the device's alarms were not working.There was no patient involvement associated with the reported event.
 
Manufacturer Narrative
H6: the device was evaluated by ventec where the reported issue of its alarms not working was confirmed.Ventec replaced the device's front panel board to resolve the reported issue.Proper device operation was then observed through functional and performance testing.The investigation determined that the cause of the reported issue was the front panel board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
2603 s goyer rd
kokomo IN 46902
Manufacturer Contact
atul sinha
22002 26th ave se
bothell, WA 98021
2067612204
MDR Report Key13691293
MDR Text Key287125028
Report Number3013095415-2022-00195
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPRT-01201-000
Device Catalogue NumberPRT-01201-000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2022
Initial Date FDA Received03/08/2022
Supplement Dates Manufacturer Received12/21/2023
Supplement Dates FDA Received01/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-