Model Number DB-1216
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2021
Event Type  malfunction  
Event Description
The returned implantable pulse generator (ipg) was analyzed and r&d investigation about the reported message had found that the log records were not properly written due to an interruption of power to the pcba (printed circuit board assembly).Before the battery is attached, the pcba receives power from the ate (automated test equipment) test fixture through an external connector.Since log record writes by firmware are asynchronous to any ate test procedures, it is possible that power is removed from the pcba during a log record write.Any corrupt log records due to this reason should not indicate any issue with the stimulator.With all the available information, boston scientific concludes that the ipg having an error message was caused by the log records not being properly written due to an interruption of power to the pcba.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Vercise Genus¿
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key13692348
MDR Text Key286750642
Report Number3006630150-2022-00887
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729985044
UDI-Public08714729985044
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/09/2023
Device Model NumberDB-1216
Device Catalogue NumberDB-1216
Device Lot Number513730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/08/2022
Initial Date FDA Received03/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-