Catalog Number 5C4482
Device Problems Fluid/Blood Leak (1250); Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that two (2) peritoneal dialysis (pd) transfer sets experienced a connection issue to the minicap which resulted in a leak.This issue was further reported that, "the set did not fit to minicap and there was excess iodine leaked out.The mini cap still ooze iodine after it is closed, even after the mini cap is changed ".This issue occurred during use of the device for peritoneal dialysis (pd) therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to b5, d9, h3, h6, and h10.Additional information for b5: clarification confirmed that only 1 unit of transfer set was used and defective (previously reported a quantity of two (2) transfer sets).H10: one (1) device was received for evaluation.A visual inspection was performed, and it was noted that the threads of the female connector were damaged.Leak testing was performed, and it was noted that there was leakage from the damaged area of the female connector.There were no issues noted with connection/disconnection to the female connector during testing.Functional testing including clear passage testing and clamp function testing were performed, and no issue were noted.The reported condition of a leak was verified.The cause of the condition could not be determined.The reported condition of a connection issue was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13694054
MDR Text Key286859281
Report Number1416980-2022-00958
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4482
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/09/2022
Initial Date FDA Received03/08/2022
Supplement Dates Manufacturer Received03/30/2022
Supplement Dates FDA Received04/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MINICAP
-
-