Model Number 2342306M
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Event Description
Information was received from a healthcare provider (hcp) via a medtronic representative regarding a patient implanted with old hardware in a revision fusion thoracolumbar with solera.It was reported that while removing the old hardware, the tip of the driver broke.There was no fragment of the driver remaining in the patient's body. there were no patient symptoms or complications as a result of this event.There was no treatment or additional surgery performed as a result of this event.The previously implanted products were no longer needed, there were no allegations against them, and they were not replaced during this event.No further complications were reported/ anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: product analysis part #2342306m ; lot# ct20g050- visual and optical examination confirmed that approximately 2mm of instrument tip has been broken and the corners of the torx tip have been rounded off.The remaining tip has also been twisted.This type of damage is consistent with torsional overload.H6: updated eval.Code method and eval.Code result post analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWEREASE¿ INSTRUMENTS
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key13707974
MDR Text Key289087164
Report Number1030489-2022-00231
Device Sequence Number1
Product Code HBE
UDI-Device Identifier00613994905727
UDI-Public00613994905727
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2342306M
Device Catalogue Number2342306M
Device Lot NumberCT20G050
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/18/2022
Initial Date FDA Received03/09/2022
Supplement Dates Manufacturer Received07/19/2022
Supplement Dates FDA Received08/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-