While nurse was touching display to do hourly numbers, the machine fell off the stand onto nurse.Nurse was able to stabilize the machine on a chair in the room and blood was returned to the patient safely.There was no injury to the nurse or the patient but therapy was interrupted.The machine is attached to the stand with three thumb nuts, two on the top and one on the bottom.All three nuts on this machine were missing and had apparently vibrated loose and fell off.Inspection of other machines found several loose or missing nuts.The machines are subject to frequent moving, including elevators, which can cause the nuts to loosen.The thumb nuts are intended to make servicing the machine easier, but don't have any kind of locking mechanism.
|