SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500 Back to Search Results
Catalog Number 10697306
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Event Description
The customer reported a discrepant low sodium result on their rp500 instrument compared to retesting of a new sample on a non siemens lab instrument.There is no report of injury due to this event.
 
Manufacturer Narrative
The customer stated that repeat testing was performed that met the clinical picture.The customer has provided instrument log files and investigation is underway.The customer did note that they use a cleaner containing a benzalkonium compound.As per the rp 500 operator's guide, quaternary ammonium compounds (such as benzalkonium) are a known interfering substance to the na sensor and should be avoided.The cause of this event is unknown.
 
Manufacturer Narrative
The customer was unable to provide the instrument log files for the dates in question.Therefore, there is insufficient information to conduct a root cause investigation for the alleged discrepant na+ result.The cause of this event is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
tim krauskopf
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key13711783
MDR Text Key296959386
Report Number3002637618-2022-00019
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
56672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10697306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received03/09/2022
Supplement Dates Manufacturer Received03/17/2022
Supplement Dates FDA Received04/11/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-