• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-XTW
Device Problems Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2022
Event Type  malfunction  
Event Description
This is filed to report material damage.It was reported that during preparation of the clip delivery system (cds), the clip was attempted to be pulled into the clip introducer (ci).However, the ci became stuck between the clip arms and the grippers.The clip was pulled off the ci and damage was observed on ci.Therefore, the cds was not used and was replaced.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
All available information was investigated, and the reported material split, cut or torn (clip introducer) was confirmed via returned device analysis.The reported difficult to advance could not be confirmed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on available information and the results of the analysis, the material split, cut or torn was related to the reported difficult to advance.The reported difficult to advance appears to be due to user technique.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13712141
MDR Text Key286879342
Report Number2024168-2022-02382
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648231001
UDI-Public08717648231001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/16/2022
Device Model NumberCDS0701-XTW
Device Catalogue NumberCDS0701-XTW
Device Lot Number10616R153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2022
Initial Date FDA Received03/09/2022
Supplement Dates Manufacturer Received03/15/2022
Supplement Dates FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-