(b)(4).The reported event is confirmed.As the sterile blister was not returned and cannot be evaluated, the sterility, or breach thereof, cannot be confirmed.Visual evaluation of the returned product found the sterile pouch exhibits creasing and holes with black debris/stain from the porous coating as a result of the implant shifting in the sterile packaging during transit.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|