• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARTICULAR SURFACE WITH HINGE POST SCREW SIZE F 12 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARTICULAR SURFACE WITH HINGE POST SCREW SIZE F 12 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Catalog Number 00588006012
Device Problems Device Dislodged or Dislocated (2923); Migration (4003)
Patient Problems Failure of Implant (1924); Joint Dislocation (2374)
Event Date 01/31/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Medical product: femoral component option for cemented use only size f right: catalog#00588001602, lot#ni; unknown nexgen tibia size 5: catalog#ni, lot#ni; unknown cement: catalog#ni, lot#ni.Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial total knee arthroplasty.Approximately six years and ten months post implantation, the patient experienced a dislocation of the knee while getting up from a chair.Subsequently, the patient underwent revision surgery during which the surgeon noted that the screw fixing the hinge had backed out.The migrated screw and articular surface was exchanged without complication.The patella was also resurfaced and a patellar component was implanted.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; b7; d4; d6a; d10; g3; h2; h4; h10.D10: medical product: femoral component option for cemented use only size f right: catalog#00588001602, lot#11015334; tibial component precoat size 5: catalog#00588000500, lot#63032091; stem extension straight 16mm dia x 100mm length(combined length 145mm): catalog#00598801016, lot#62955353; stem extension straight 10mm dia x 100mm length(combined length 145mm): catalog#00598801010, lot#62937538; palacos r-40g cement: catalog#66017750, lot#81104418, qty#2.The investigation is in process.Once the investigation is complete, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial right total knee arthroplasty.Approximately six years and four months post implantation, the patient experienced a dislocation of the knee while getting up from a chair.Subsequently, the patient underwent revision surgery during which the surgeon noted that the screw fixing the hinge had backed out.The migrated screw and articular surface were exchanged without complication.The patella was also resurfaced and a patellar component was implanted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d9; g3; h2; h3; h6; h10.Visual examination of the returned products identified that the articular surface shows signs of being implanted (gouged, nicked, wear).The returned hinge post extension component showed signs of being implanted; damage to the threads and striations on the shaft.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: revision records: loosening/disassociation of the screw without trauma, no signs of implant loosening, no hinge fracture, no damage to poly identified, poly exchange and patellar resurfacing performed without complication.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information at time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE WITH HINGE POST SCREW SIZE F 12 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13712763
MDR Text Key286876099
Report Number0001822565-2022-00645
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K013385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Catalogue Number00588006012
Device Lot Number62996374
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/11/2022
Initial Date FDA Received03/09/2022
Supplement Dates Manufacturer Received03/17/2022
06/17/2022
Supplement Dates FDA Received03/23/2022
06/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.; SEE H10 NARRATIVE.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
Patient Weight85 KG
-
-