• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS U 701 MICROSCOPY ANALYZER; AUTOMATED URINALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS U 701 MICROSCOPY ANALYZER; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Model Number U701
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation determined the issue was consistent with poor sample quality.The quality of the sample caused the analyzer to not detect any cells due to blurriness and mucus.The investigation could not identify a product problem.
 
Event Description
The initial reporter stated they received false negative results for one patient sample tested for bacteria and leukocytes on a cobas u 701 microscopy analyzer.The leukocyte and bacteria results measured from the sample were negative.The sample was found to be completely full of leukocytes and bacteria.The pictures from the u 701 analyzer were noted to be very fuzzy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS U 701 MICROSCOPY ANALYZER
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13713724
MDR Text Key293077655
Report Number1823260-2022-00636
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
ASKU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberU701
Device Catalogue Number06390501001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2022
Initial Date FDA Received03/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-