RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
The customer reported internal leak, patient disconnect.The customer reported that the unit was in use on patient , but there was no patient harm reported.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The field service engineer (fse) reported that the customer complaint is unit is having issues with delivering pressure.The fse and onsite hospital staff were not able to duplicate this condition.The fse was not able to locate any failing diagnostic codes.The fse ran the unit in normal ventilation and unit operated correctly.Per service manual fse completed a full performance verification test (pvt) test and all tests passed.As a precaution the fse tested the pressure and air flows an additional time and units operated within tolerances.The fse received approval from hospital staff that the unit was working correctly at the time of service.No problem found with the unit.
|
|