• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-67A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2022
Event Type  Injury  
Manufacturer Narrative
This mdr is result of a retrospective review of complaints.User was explained that having two sensors near to each other in the same arm could cause signal interference preventing the proper system use.User was advised discuss the situation with their health care professional, including removing both sensors and to get inserted with another one in another pocket, preferably in the other arm.No further investigation is necessary for this complaint.
 
Event Description
On 28 january 2022,senseonics was made aware of an incident where user reported having 2 sensors inserted on the same arm.User was explained that having two sensors near to each other in the same arm could cause signal interference preventing the proper system use.User was advised discuss the situation with their hcp, including removing both sensors and to get inserted with another one in another pocket, preferably in the other arm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key13734710
MDR Text Key287018295
Report Number3009862700-2022-00044
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491022349
UDI-Public817491022349
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/23/2022
Device Model Number102096-67A
Device Catalogue NumberFG-4200-00-301
Device Lot NumberWP08507
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/28/2022
Initial Date FDA Received03/11/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexFemale
-
-