Device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection results: observation of the returned breathing circuit revealed no abnormalities such as damage related to the reported event.Functional testing: we conducted a leak test on the breathing circuit, but no leaks were confirmed, and it was confirmed that the product conformed to the standard.The reported event was not confirmed.The cause of the reported problem could not be determined.
|