RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/25/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported that the unit has internal leak.The customer reported that the unit was not in use on patient.Per biomed the issue was discovered during setup.The customer contacted product support and requested for service repair.The biomed reported that he replaced the solenoid 4 valve to address the reported based on information provided and/or service performed, the customer's alleged malfunction was confirmed, or failed to meet specifications.The cause of the reported issue is solenoid 4 leaking.
|
|