Catalog Number 8065752914 |
Device Problems
Overheating of Device (1437); Defective Component (2292); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A physician reported that an ophthalmic handpiece was getting heat and lost efficiency.There was no patient harm.Additional information has been requested, but none received till date.
|
|
Manufacturer Narrative
|
The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
Additional information was received and indicated that handpiece was not recognized by console.
|
|
Manufacturer Narrative
|
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, product was returned for evaluation.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The phacoemulsification handpiece (hp) was received for evaluation.A visual assessment of the returned sample revealed no obvious nonconformity.The returned sample was connected to a calibrated system.The handpiece tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.The handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the handpiece to meet specifications the handpiece was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|