• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 102956
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2022
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the centrimag motor had issues with centering the rotor.The motor would not connect properly to the centrimag console.Additional information was received that the cause of the issue was not identified and the issue did not resolve.The motor was removed from use and returned for analysis.There were no patient consequences.Related mfr.Report number: 3003306248-2022-00042.
 
Manufacturer Narrative
G2: health professional was selected inadvertently.Manufacturer's investigation conclusion: the reported event of the motor not connecting properly to the console was confirmed.The centrimag motor was returned for analysis to the european distribution center (edc).When connected to a test system a ¿motor disconnected: m2¿ alarm was active.The motor cable was tested, and an open lead was found.The value did not fluctuate when the motor cable was manipulated.The motor was scrapped; however, the customer requested the motor be sent back.The root cause for the reported event was conclusively determined to be wire fatigue within the cable.The device history records were reviewed for the centrimag motor and the motor was found to pass all manufacturing and qa specifications.Centrimag motor instructions for use instructs to always have a back-up centrimag motor available.Centrimag blood pump instructions for use states "always have a spare centrimag blood pump, back-up console and equipment available for change out." centrimag primary console operating manual warns "one back-up console and motor are required in the immediate vicinity of each patient whenever the centrimag blood pump is used.The back-up console must be connected to the back-up motor, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the primary console or primary motor experience a malfunction." no further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, OUS
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key13759776
MDR Text Key287709797
Report Number3003306248-2022-00041
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140078
UDI-Public7640135140078
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2022
Initial Date FDA Received03/14/2022
Supplement Dates Manufacturer Received03/30/2022
Supplement Dates FDA Received04/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberFA-Q318-MCS-1
Patient Sequence Number1
-
-