G2: health professional was selected inadvertently.Manufacturer's investigation conclusion: the reported event of the motor not connecting properly to the console was confirmed.The centrimag motor was returned for analysis to the european distribution center (edc).When connected to a test system a ¿motor disconnected: m2¿ alarm was active.The motor cable was tested, and an open lead was found.The value did not fluctuate when the motor cable was manipulated.The motor was scrapped; however, the customer requested the motor be sent back.The root cause for the reported event was conclusively determined to be wire fatigue within the cable.The device history records were reviewed for the centrimag motor and the motor was found to pass all manufacturing and qa specifications.Centrimag motor instructions for use instructs to always have a back-up centrimag motor available.Centrimag blood pump instructions for use states "always have a spare centrimag blood pump, back-up console and equipment available for change out." centrimag primary console operating manual warns "one back-up console and motor are required in the immediate vicinity of each patient whenever the centrimag blood pump is used.The back-up console must be connected to the back-up motor, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the primary console or primary motor experience a malfunction." no further information was provided.The manufacturer is closing the file on this event.
|