Model Number PROLACTIN G2 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/04/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Na.
|
|
Event Description
|
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys prolactin gen.2 assay ver.2.1 on a cobas e 411 immunoassay analyzer.The sample initially resulted in a prolactin value of 42.22 ng/ml and this value was reported outside of the laboratory.The sample was sent to another laboratory for testing using a chemiluminescence method and when tested with this method, the prolactin result was 22.50 ng/ml.The sample was treated with polyethylene glycol (peg) and repeated with the chemiluminescence method, resulting in a value of 19.92 ng/ml (89 % recovery).The e411 analyzer serial number is (b)(4).
|
|
Manufacturer Narrative
|
Calibration signals were within expectations.Quality controls were within specified ranges for the dates of sample measurements.The sample coagulation time was shorter than recommended by tube manufacturers.Siemens centaur has a lower macro prolactin detection rate than the elecsys assay.Therefore, it is necessary to perform polyethylene glycol (peg) precipitation for the elecsys prolactin assay which requires this measure for "unexpected high prolactin" samples.Per product labeling: "in case of implausible high prolactin values a precipitation by polyethylene glycol (peg) is recommended in order to estimate the amount of the biological active monomeric prolactin." the investigation could not identify a product problem.
|
|
Search Alerts/Recalls
|