• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problems Crack (1135); Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6) via a fisher and paykel healthcare (f&p) field representative, that a mr290v vented autofeed humidification chamber was found cracked prior patient use.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected.Results: visual inspection of the returned mr290 chamber revealed vertical crack lines and stress marks were observed on the chamber dome.Conclusion: we are unable to determine the cause of the reported event.However, our investigation indicates that the damage was most likely due to transport damage.Every mr290 chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290 chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290hfv vented autofeed humidification chamber state the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers."."perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient."."ensure appropriate ventilator and flow source alarms are set before connecting breathing set to patient."."use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
 
Event Description
A distributor reported on behalf of a healthcare facility in japan via a fisher and paykel healthcare (f&p) field representative, that a mr290v vented autofeed humidification chamber was found cracked prior patient use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology dr.
suite 100
irvine, CA 92618
9494534000
MDR Report Key13766570
MDR Text Key291635557
Report Number9611451-2022-00240
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number210625
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/04/2022
Initial Date FDA Received03/15/2022
Supplement Dates Manufacturer Received04/12/2022
Supplement Dates FDA Received04/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-