Model Number 10618 |
Device Problems
Entrapment of Device (1212); Difficult to Advance (2920); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/26/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that device entrapment and stent damage occurred.The patient was diagnosed with small vessel disease (svd) of the left circumflex (lcx) artery.Following pre-dilatation, a 2.25 x 16 synergy drug eluting stent was loaded onto the guidewire.However, during introduction, the stent got stuck on the guidewire outside the patient and the stent was damaged.The physician removed the device by cutting the guidewire.The procedure was completed with another of the same device and a new guidewire.No patient complications were reported and the patient condition post procedure was stable.
|
|
Event Description
|
It was reported that stent got stuck on guidewire occurred.The patient was diagnosed with small vessel disease (svd) of the left circumflex (lcx) artery.Following pre-dilatation, a 2.25 x 16 synergy drug eluting stent was loaded onto the guidewire.However, during introduction, the stent got stuck on the guidewire outside the patient and the stent was damaged.The physician removed the device by cutting the guidewire.The procedure was completed with another of the same device and a new guidewire.No patient complications were reported and the patient condition post procedure was stable.
|
|
Manufacturer Narrative
|
The synergy ous mr 2.25 x 16mm stent delivery system (sds) was returned for analysis.The device was returned with a segment of customer guidewire inserted.The guidewire diameter measured 0.014.This segment of guidewire was removed without issue.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent od (outer diameter) was measured within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip identified tip damage.A visual and tactile examination of the hypotube identified no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen identified a 9.5 cm length of stretching on the polymer shaft distal to the exchange port, a polymer shaft kink at 7 cm distal to the exchange port, and bunching on the polymer shaft inner at 5.5 cm proximal to the distal tip.The device could not be loaded or tracked through a 0.014 guidewire due to the extent of the polymer shaft damage.
|
|
Search Alerts/Recalls
|
|