Model Number 16-02-85
Device Problem Calibration Problem (2890)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative recalibrated the temperature sensor on the patient tank and the issue could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The customer reported that an error code was displayed as soon as patient heating started but the customer did not recall the exact error code.During service activity for this complaint the technician found an error code associated to the stirrer motor (a separated report has been submitted for this issue) and no other error occurred.However, it cannot be confirmed if the error experienced by the customer was due to the stirrer motor or to the temperature sensors issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that displayed temperatures on a heater-cooler system 3t were off by about 1.5 degree celsius.There was no report of patient injury.
 
Manufacturer Narrative
H10: a device service history review has been performed and identified that the unit was manufactured in 2011 and no other similar events have been reported.Based on service activity findings and considering the manufacturing date of involved unit, it cannot be ruled out that reported failure was due to temperature sensor loosing its initial adjustment due to a calibration that reasonably drifted over time.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key13768247
MDR Text Key287343116
Report Number9611109-2022-00118
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Followup
Report Date 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/15/2022
Supplement Dates Manufacturer Received04/14/2022
Supplement Dates FDA Received05/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-