Roche received complaints from customers reporting the generation of false mutation detected results for the exon 20 insertion (ex20ins) mutation when using the cobas® egfr mutation test v2.During in-house testing using customer-provided ffpet samples, an ex20ins false mutation detected result was reproduced for one out of 8 ffpet samples, which was processed following the validated sample preparation method from the instructions for use.Although the majority of cases reported were from users using ffpet samples, the generation of false mutation detected ex20ins results with plasma specimens was reported in one case.A false mutation detected ex20ins result could lead to harm under specific scenarios.Consignees will be notified of the issue with instruction to follow the instructions for use for sample input requirements.Additionally, if an ex20ins mutation detected result is generated with the cobas® egfr mutation test v2, customers must confirm the result with another method (e.G., sequencing or other pcr-based tests).The most likely cause of the discrepancy observed is due to false positive detection of ex20ins.It is also possible that the use of a sample type not indicated in the product's method sheets is contributing to the issue.Another possibility is that the sample contains low levels of mutation near the cut-off of the assay.Samples near the lod of the assay are expected to waver between positive and negative results with repeat testing.(b)(4).
|