DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery run test.Specifically, the battery runtime was 95 minutes and below the specification of 135 minutes.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: the getinge field service engineer (fse) that encountered the issue replaced the 12v batteries (0146-00-0039) and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.(b)(6).
|
|
Manufacturer Narrative
|
Updated fields; b4, g3, g6, g7, h2, h4, h6, h10, h11 corrected fields: g1.
|
|
Event Description
|
N/a.
|
|