Model Number N/A
Device Problems Break (1069); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).The client has indicated that the product is in process of being returned to zimmer biomet for an investigation.The investigation is currently underway.Once the investigation has concluded, a follow-up mdr will be submitted.
 
Event Description
It was reported, that: the surgeon was trialling and when she removed it broke.She found all the pieces and made sure no pieces were left, inpatient.Delay: 5 min.Patient outcome: prolonged surgery.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.Visual examination of the returned product identified the trial bearing has two fractures.Both are from the slots on either side of the bearing.The bearing appears well used and has potentially been in the field for approximately 3 years and 8 months.A review of the device history records did not identify any discrepancies that would have contributed to the reported event.The device is used for treatment.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF TRL BRG W/SLOTS SML 3MM
Type of Device
BMET GENERAL KNEE INSTRUMENTS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13789249
MDR Text Key289725645
Report Number3002806535-2022-00103
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279346608
UDI-Public05019279346608
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-422698
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2022
Initial Date FDA Received03/16/2022
Supplement Dates Manufacturer Received07/14/2022
Supplement Dates FDA Received07/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-