• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M Back to Search Results
Model Number BI70000027230
Device Problems Arcing (2583); Temperature Problem (3022); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2022
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the x-ray charge 1 board and motion charge board and tray2, tray3 batteries was damaged.The burn smell was from charge board.B01, c02, d02 applicable codes.Information references the main component of the system.Other relevant device(s) are: product id: bi71000526, serial/lot #: unknown, ubd: , udi#: ; product id: bi71000525, serial/lot #: unknown, ubd: , udi#: ; product id: bi71000581, serial/lot #: unknown, ubd: , udi#: ; product id: bi71000125, serial/lot #: none, ubd: , udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the device has burnt smell after power failure. after boot up the doctor has seen a spark under the ias system, then system shutdown itself.They plugged off the power cable after that.There was no patient present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13790071
MDR Text Key290100862
Report Number3004785967-2022-00188
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K173664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000027230
Device Catalogue NumberBI70000027230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received03/16/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-