• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG TC DE'BAKEY NDL HOLDER DEL SERR 180MM; DUROGRIP NEEDLEHOLDERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG TC DE'BAKEY NDL HOLDER DEL SERR 180MM; DUROGRIP NEEDLEHOLDERS Back to Search Results
Model Number BM035R
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2021
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap (b)(4) that a tc de'bakey ndl holder del serr 180mm (part # bm035r) was used during a cardiac surgery performed on (b)(6) 2021.According to the complainant, during the surgery, the needle holder broke.Reportedly, in the middle of the procedure, a part of the tungsten came off the mouth of the needle holder and detached into the patient.The fragment was retrieved from the patient.The complaint device was not returned to the manufacturer for evaluation.An additional medical intervention was necessary.Although requested, additional information has not been made available.The adverse event is filed under aesculap (b)(4) reference (b)(4).
 
Manufacturer Narrative
H6 - codes updated.Investigation: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.According to the investigation results, a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TC DE'BAKEY NDL HOLDER DEL SERR 180MM
Type of Device
DUROGRIP NEEDLEHOLDERS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key13792936
MDR Text Key288620094
Report Number9610612-2022-00057
Device Sequence Number1
Product Code HXK
UDI-Device Identifier04038653028092
UDI-Public4038653028092
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBM035R
Device Catalogue NumberBM035R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2022
Initial Date FDA Received03/17/2022
Supplement Dates Manufacturer Received03/31/2022
Supplement Dates FDA Received04/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-