Based on the investigation analysis, there was temporary mismatch between the sensor readings and the fingerstick measurement.The hypo alert was asserted incorrectly at the time of the reported event due to the temporary difference between the sensor reading and the fingerstick measurement.After the event, the system displayed better agreement between the sensor readings and fingerstick measurements.As per notes, user wasn't symptomatic and didn't seek medical treatment for her hypoglycemic event.User was advised to let the hcp know of these events.H3.Device evaluated by manufacturer? yes.Not returned to manufacturer.H6.Investigation findings updated to 213.H6.Investigation conclusions updated to 67.
|